The quantum efficiency (QE) is unsuitable to characterise the performance of an imaging detector. Instead, the detective quantum efficiency (DQE), which indicates how much the detection process deteriorates the signal-to-noise ratio of the input image, is a more convenient parameter. Nevertheless QE plays a relevant role in the DQE formulation, as we have shown in the present paper, which is intended as a continuation and improvement of a previous paper of the same authors. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 381
- Journal Issue
- 1
- Journal Page Range
- p. 157-160.
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 28014401
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- FLUCTUATIONS; POSITION SENSITIVE DETECTORS; QUANTUM EFFICIENCY; RADIATION DETECTION; SIGNAL-TO-NOISE RATIO
- Descriptors DEC
- DETECTION; EFFICIENCY; MEASURING INSTRUMENTS; RADIATION DETECTORS; VARIATIONS