Published October 21, 1996 | Version v1
Journal article

The role of the quantum efficiency on the DQE of an imaging detector

  • 1. Padua Univ. (Italy). Dipt. di Fisica

Description

The quantum efficiency (QE) is unsuitable to characterise the performance of an imaging detector. Instead, the detective quantum efficiency (DQE), which indicates how much the detection process deteriorates the signal-to-noise ratio of the input image, is a more convenient parameter. Nevertheless QE plays a relevant role in the DQE formulation, as we have shown in the present paper, which is intended as a continuation and improvement of a previous paper of the same authors. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
381
Journal Issue
1
Journal Page Range
p. 157-160.
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
Netherlands
INIS RN
28014401
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
FLUCTUATIONS; POSITION SENSITIVE DETECTORS; QUANTUM EFFICIENCY; RADIATION DETECTION; SIGNAL-TO-NOISE RATIO
Descriptors DEC
DETECTION; EFFICIENCY; MEASURING INSTRUMENTS; RADIATION DETECTORS; VARIATIONS