Published July 2013 | Version v1
Journal article

Development of a hotspot detector with an acrylic filter and dose rate survey meters

  • 1. National Inst. of Radiological Sciences, Chiba, Chiba (Japan)
  • 2. Hitachi-Aloka Medical, Ltd., Oume, Tokyo (Japan)

Description

Fukushima and adjacent regions still have a large number of high dose rate areas called hotspots. It is necessary to know these hotspots for efficient decontamination of radioactive substances such as 137Cs and for relief of residents coming home. To find the hotspots rapidly, we have to specify the direction of the area where the dose rate is at least 1μSv/h higher than those of surroundings. We have developed a detector that consists of an acrylic filter and three NaI(Tl) scintillation survey meters, and the detector can be expected to indicate the direction of the hotspot in the short time. A basic performance of the detector was examined by using acrylic filters of 10, 15, 20 and 25cm diameter and a tiny sealed 137Cs source of 3 MBq as the alternative of a hotspot. It demonstrated the possibility of identifying the direction of γ-rays emitted from the source in 90 seconds. (author)

Availability note (English)

Available from http://dx.doi.org/10.3769/radioisotopes.62.439

Additional details

Identifiers

Publishing Information

Journal Title
Radioisotopes (Tokyo)
Journal Volume
62
Journal Issue
7
Journal Page Range
p. 439-445
ISSN
0033-8303

Optional Information

Notes
8 refs., 8 figs.; This record replaces 45006457