Description

The present studies deal with the optimum physical conditions which seriously affect the neutron activation analysis technique efficiency. An experimental work for the efficiency calibration of hyper pure germanium detectors especially for environmental studies is presented. This work showed that the tested parameters, under consideration, distance, mass and measured time, reveal a significant effect on the obtained data. These results, intern, affect the accuracy of the measurements. Further work on the test of other parameters is planned in our laboratory using special treatments and applying special computer programs