Published May 10, 2006 | Version v1
Journal article

The use of energy windowing to discriminate SNM from NORM in radiation portal monitors

  • 1. Pacific Northwest National Laboratory, MS P8-20, P.O. Box 999, Richland, WA 99352 (United States)

Description

Energy windowing is an algorithmic alarm method that can be applied to plastic scintillator-based radiation portal monitor (RPM) systems to improve operational sensitivity to certain threat sources while reducing the alarm rates from naturally occurring radioactive material. Various implementations of energy windowing have been tested and documented by industry and at Pacific Northwest National Laboratory, and are available in commercial RPMs built by several manufacturers. Moreover, energy windowing is being used in many deployed RPMs to reduce nuisance alarms and improve operational sensitivity during the screening of cargo. This paper describes energy windowing algorithms and demonstrates how these algorithms succeed when applied to 'controlled' experimental measurements and 'real world' vehicle traffic data

Additional details

Identifiers

DOI
10.1016/j.nima.2006.01.053;
PII
S0168-9002(06)00105-7;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
560
Journal Issue
2
Journal Page Range
p. 373-387
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
37094186
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ALGORITHMS; CARGO; MANUFACTURERS; MONITORS; PLASTIC SCINTILLATORS; RADIATION DETECTION; RADIOACTIVE MATERIALS; SECURITY; SENSITIVITY; SHIELDING; VEHICLES
Descriptors DEC
DETECTION; MATERIALS; MATHEMATICAL LOGIC; MEASURING INSTRUMENTS; PHOSPHORS