Coherent-Compton scattering for the assessment of bone mineral content using heavily filtered x-ray beams
Creators
- 1. Royal United Hospital, Bath (UK). Regional Medical Physics Dept.
Description
The feasibility of using heavily filtered x-ray beams to assess trabecular bone mineral content has been investigated by measuring the ratio of coherent to incoherent scattered x-rays with a high purity Ge detector. The technique uses the strong dependence of coherent scattering on the effective atomic number of the scattering medium. With an x-ray beam filtered with a high-atomic-number filter, a spectrum characterised by a sharp discontinuity at the K-absorption edge is produced. Analysis of the spectral shape after scattering allows the coherent to Compton scattering ratio to be obtained. Theoretical and experimental results from phantom studies are presented and a comparison made between the results obtained with x-ray beams and radionuclide sources respectively. The influence of overlying tissue thickness on the sensitivity of the measurements is demonstrated. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 30
- Journal Issue
- 6
- Series
- Phys. Med. Biol.
- Journal Page Range
- 531-539
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16064135
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ATOMIC NUMBER; COHERENT SCATTERING; COMPTON EFFECT; FILTERS; GE SEMICONDUCTOR DETECTORS; K ABSORPTION; MINERALS; OSTEOPOROSIS; PHANTOMS; PHOTON BEAMS; TRABECULAR BONE; X RADIATION
- Descriptors DEC
- ABSORPTION; BASIC INTERACTIONS; BEAMS; BODY; BONE TISSUES; CONNECTIVE TISSUE; DISEASES; ELASTIC SCATTERING; ELECTROMAGNETIC INTERACTIONS; ELECTROMAGNETIC RADIATION; INTERACTIONS; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MOCKUP; RADIATION DETECTORS; RADIATIONS; SCATTERING; SEMICONDUCTOR DETECTORS; SKELETAL DISEASES; STRUCTURAL MODELS; TISSUES