Published February 1987
| Version v1
Journal article
Performance characteristics studies of image intensifier having short luminescence decay time
Creators
- 1. Lawrence Berkeley Lab., Univ. of California, Berkeley, CA 94720
Description
The characteristics of a modified first generation 40 mm-diameter VARO 1248-1 image intensifier which uses a very short luminescence decay time phosphor screen (P-46) have been investigated. Some typical intensifier characteristics - such as luminous gain and spatial resolution - are compared with data provided by the manufacturer for nonmodified device. Characteristics which are not generally available from manufacturer were also measured. These include the phosphor spectral emission characteristics and luminescence decay time. The measuring techniques used and description of measuring systems are given
Additional details
Publishing Information
- Journal Title
- IEEE Trans. Nucl. Sci.
- Journal Volume
- NS-34
- Journal Issue
- 1
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 453-457
- ISSN
- 0018-9499
- CODEN
- IETNA
Conference
- Title
- Nuclear science and nuclear power systems symposium.
- Dates
- 29-31 Oct 1986.
- Place
- Washington, DC (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18061997
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DATA PROCESSING; DECAY; EMISSION; GAIN; IMAGE INTENSIFIERS; LUMINESCENCE; MEASURING METHODS; PERFORMANCE TESTING; PHOSPHORS; SPATIAL RESOLUTION; TIME DEPENDENCE
- Descriptors DEC
- AMPLIFICATION; PHOTON EMISSION; RESOLUTION; TESTING
Optional Information
- Secondary number(s)
- CONF-861007--.