A novel dose uncertainty model and its application for dose verification
Creators
- 1. Department of Radiation Oncology, University of Florida, Gainesville, Florida 32610 (United States)
- 2. Department of Biomedical Engineering, Catholic Medical University (Korea, Republic of)
- 3. Department of Nuclear and Radiological Engineering, University of Florida, Gainesville, Florida 32610 (United States)
Description
Based on statistical approach, a novel dose uncertainty model was introduced considering both nonspatial and spatial dose deviations. Non-space-oriented uncertainty is mainly caused by dosimetric uncertainties, and space-oriented dose uncertainty is the uncertainty caused by all spatial displacements. Assuming these two parts are independent, dose difference between measurement and calculation is a linear combination of nonspatial and spatial dose uncertainties. Two assumptions were made: (1) the relative standard deviation of nonspatial dose uncertainty is inversely proportional to the dose standard deviation σ, and (2) the spatial dose uncertainty is proportional to the gradient of dose. The total dose uncertainty is a quadratic sum of the nonspatial and spatial uncertainties. The uncertainty model provides the tolerance dose bound for comparison between calculation and measurement. In the statistical uncertainty model based on a Gaussian distribution, a confidence level of 3σ theoretically confines 99.74% of measurements within the bound. By setting the confidence limit, the tolerance bound for dose comparison can be made analogous to that of existing dose comparison methods (e.g., a composite distribution analysis, a γ test, a χ evaluation, and a normalized agreement test method). However, the model considers the inherent dose uncertainty characteristics of the test points by taking into account the space-specific history of dose accumulation, while the previous methods apply a single tolerance criterion to the points, although dose uncertainty at each point is significantly different from others. Three types of one-dimensional test dose distributions (a single large field, a composite flat field made by two identical beams, and three-beam intensity-modulated fields) were made to verify the robustness of the model. For each test distribution, the dose bound predicted by the uncertainty model was compared with simulated measurements. The simulated measurements were within the tolerance bound as expected by a statistical prediction of the model. Using the dose uncertainty distributions, an uncertainty length (uncertainty area and uncertainty volume for two-dimensional and three-dimensional, respectively) histogram (a plot of the dose uncertainty of 1σ received by a length of field) was made. The histogram provides additional information on superiority of a treatment plan in terms of uncertainty. In summary, the uncertainty model provides the dose comparison tool as well as the evaluation tool of a treatment planning system
Additional details
Identifiers
- DOI
- 10.1118/1.1924329;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 32
- Journal Issue
- 6
- Journal Page Range
- p. 1747-1756
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37037516
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; DOSIMETRY; GAUSS FUNCTION; MODULATION; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; TOLERANCE; VERIFICATION
- Descriptors DEC
- DOSES; EVALUATION; FUNCTIONS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY
Optional Information
- Notes
- (c) 2005 American Association of Physicists in Medicine