Published May 2016 | Version v1
Journal article

The segmented cylindrical mirror analyzer (CMA)

Creators

Description

Highlights: • A method is presented applying segmentation to the electrodes of cylindrical mirror analyzer. • Improvement was found in resolution over the standard CMA by a factor of ∼10. • Segmentation of both outer and inner electrodes was required. - Abstract: The standard cylindrical mirror analyzer (CMA) has been modified by dividing the inner and outer electrodes into segments and setting the voltages on these segments to obtain the best value for the energy resolution. It has been found that dividing both the inner and outer electrodes using only one segment has resulted in an energy resolution of 0.017% for entrance angles of 42.3 ± 6°. This corresponds to more than an order of magnitude improvement in the resolution over that of the standard instrument.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.elspec.2016.02.004

Additional details

Identifiers

DOI
10.1016/j.elspec.2016.02.004;
PII
S0368-2048(16)30010-X;

Publishing Information

Journal Title
Journal of Electron Spectroscopy and Related Phenomena
Journal Volume
209
Journal Page Range
p. 46-52
ISSN
0368-2048
CODEN
JESRAW

Optional Information

Copyright