Published October 1981 | Version v1
Report Restricted

Measurement of wiretension in drift tubes by resonance in a magnetic field

Description

Wiretension in drift chambers is commonly measured by observation of resonance frequency. In drift tubes, where the wire cannot be seen by eye, a vibration of the wire has to be detected in other ways. The method described here uses a simple bridge circuit for measuring the change in dc-resistance of the resonating wire. The decrease of the resistance (0.3%) is due to self-cooling (0.50C) by convection when the wire starts swinging. (orig.)

Availability note (English)

MF available from INIS under the Report Number.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
4 p.
Report number
DESY--81-065