Published January 1992 | Version v1
Journal article

Evaluation of a new VUV/soft x-ray toroidal grating monochromator with a movable exit slit

  • 1. Synchrotron Radiation Center, University of Wisconsin Madison, Stoughton, Wisconsin 53589 (United States)

Description

The performance of a toroidal grating monochromator can be significantly improved by compensating for the wavelength dependent defocusing aberration by means of a movable exit slit. Model calculations of a movable exit slit design show a resolving power that varies nearly linearly with wavelength over the scan range. The energy resolution follows the slope between two optimized resolution minima of a fixed slit TGM. The overall improvement in resolution due to the movable slit depends critically on the figure error of the toroid surface. The Al(LII,III) core edges provide a means to calibrate gratings and determine the resolution at 72.8 eV. A method is described for using the shift in core edge with exit slit position to enable final alignment of the exit slit motion providing stability of energy calibration of 2 meV over the scan range. Comparison of calculations and measured resolution suggests that slope errors remain the limiting factor in performance

Additional details

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
63
Journal Issue
1
Series
Rev. Sci. Instrum.
Journal Page Range
1269-1273
ISSN
0034-6748
CODEN
RSINA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
23059031
Subject category
S43: PARTICLE ACCELERATORS;
Descriptors DEI
DESIGN; MONOCHROMATORS; PERFORMANCE; SOFT X RADIATION; SYNCHROTRON RADIATION SOURCES; ULTRAVIOLET RADIATION
Descriptors DEC
ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; RADIATION SOURCES; RADIATIONS; X RADIATION