Investigation of defects in CVD diamond: Influence for radiotherapy applications
- 1. LIST - CEA-Recherche Technologique, DETECS/SSTM/LTD-CEA/Saclay, 91191 Gif-sur-Yvette (France)
- 2. LIST- CEA-Recherche Technologique, DETECS/SSTM/LTD-CEA/Saclay, 91191 Gif-sur-Yvette (France)
- 3. European synchrotron Radiation Facility, Grenoble (France)
Description
In this study we present the potentialities of CVD diamond as an ionisation chamber for radiotherapy applications. Trapping levels present in CVD diamond are characterised using Thermally Stimulated Current (TSC) method with X-ray sources. The influence of the corresponding defects on the detector response is investigated and compared to those observed in natural diamond. Also, their spatial distribution across a large area polycrystalline diamond ionisation chamber is discussed. Results show the relative influence of two different populations of trapping levels in CVD diamond whose effect is crucial for radiotherapy applications. To partially overcome the defect detrimental effects, we propose to use CVD diamond ionisation chambers at moderate temperatures from 70 to 100 deg. C that could be provided by self heating of the device, for a dramatically improved stability and reproducibility
Additional details
Identifiers
- DOI
- 10.1016/j.nima.2005.06.015;
- PII
- S0168-9002(05)01179-4;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 552
- Journal Issue
- 1-2
- Journal Page Range
- p. 105-111
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 5. international conference on radiation effects on semiconductor materials, detectors and devices
- Dates
- 10-13 Oct 2004
- Place
- Florence (Italy)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37045014
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CHEMICAL VAPOR DEPOSITION; DEFECTS; DIAMONDS; IONIZATION CHAMBERS; POLYCRYSTALS; RADIOTHERAPY; SPATIAL DISTRIBUTION; THERAPEUTIC USES; TRAPPING; X-RAY SOURCES
- Descriptors DEC
- CARBON; CHEMICAL COATING; CRYSTALS; DEPOSITION; DISTRIBUTION; ELEMENTS; MEASURING INSTRUMENTS; MEDICINE; MINERALS; NONMETALS; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIATION SOURCES; RADIOLOGY; SURFACE COATING; THERAPY; USES
Optional Information
- Copyright
- Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.