Analysis of electron beam output factors by Monte Carlo method

  • 1. Department of Radiation Oncology, Fourth Hospital of Hebei Medical University, Shijiazhuang (China)
  • 2. Cangzhou People's Hospital, Cangzhou (China)
  • 3. Hebei Children's Hospital, Shijiazhuang (China)

Description

Objective: To investigate the application of the Monte Carlo dose calculation of output factors for electron beams in radiotherapy. Methods: The code EGS4/MCTP was used to simulate the head of a medical linear accelerator (Varian 23EX) to calculate the output factors for 6 MeV, 9 MeV and 18 MeV electron beams. The source-to-surface distance used was 100 cm. The field sizes ranged from 2 cm × 2 cm to 25 cm × 25 cm. The calculated output factors agreed with the corresponding measured factors which were measured by the IBA Phantom system to within 2%. Then, the output factors of direct articles and indirect articles which were under different energy and various cone-insert combinations were calculated by the code EGS4/MCTP. Results: The calculated output factors agreement with the measurements is found to be mostly under the 1% level. The variation of output factors depends on the characteristics of the beams and the modifications that the various cone-insert combinations introduce to these characteristics. Conclusions: Monte Carlo dose calculations for electron beams in homogeneous water phantoms have been demonstrated to be accurate under the 1% level in comparison with measurements. The output factors were influenced by energy and cone-insert combinations in complex ways. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiation Oncology
Journal Volume
23
Journal Issue
3
Journal Page Range
p. 265-268
ISSN
1004-4221

Optional Information

Notes
2 tabs., 17 refs.; http://dx.doi.org/10.3760/cma.j.issn.1004-4221.2014.03.023