Published March 1992 | Version v1
Journal article

Study of the decay of 131I and 134Cs

  • 1. China Inst. of Atomic Energy, Beijing (China)

Description

Relative emission probabilities of gamma-rays from the decay of 131I and 134Cs are measured with a calibrated coaxial Ge detector. Experimental uncertainties are about ±1% (1 σ) for strong gamma-rays. These two nuclides are useful for the calibration of Ge detector in 80-1400 keV

Additional details

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
26
Journal Issue
3
Journal Page Range
p. 75-79.
ISSN
1000-6931
CODEN
YKJIEZ