Sensitivity and accuracy of volumetry of pulmonary nodules on low-dose 16- and 64-row multi-detector CT: an anthropomorphic phantom study
Creators
- 1. University of Groningen, University Medical Center Groningen, Center for Medical Imaging-North East Netherlands, Department of Radiology, Groningen (Netherlands)
- 2. University of Groningen, University Medical Center Groningen, Department of Radiology, EB44, P.O. Box 30.001, Groningen (Netherlands)
- 3. University Medical Center Utrecht, Department of Radiology, Utrecht (Netherlands)
- 4. University of Groningen, University Medical Center Groningen, Department of Epidemiology, Groningen (Netherlands)
Description
To assess the sensitivity of detection and accuracy of volumetry by manual and semi-automated quantification of artificial pulmonary nodules in an anthropomorphic thoracic phantom on low-dose CT. Fifteen artificial spherical nodules (diameter 3, 5, 8, 10 and 12 mm; CT densities -800, -630 and +100 HU) were randomly placed inside an anthropomorphic thoracic phantom. The phantom was examined on 16- and 64-row multidetector CT with a low-dose protocol. Two independent blinded observers screened for pulmonary nodules. Nodule diameter was measured manually, and volume calculated. For solid nodules (+100 HU), diameter and volume were also evaluated by semi-automated software. Differences in observed volumes between the manual and semi-automated method were evaluated by a t-test. Sensitivity was 100 % for all nodules of >5 mm and larger, 60-80 % for solid and 0-20 % for non-solid 3-mm nodules. No false-positive nodules but high inter-observer reliability and inter-technique correlation were found. Volume was underestimated manually by 24.1 ± 14.0 % for nodules of any density, and 26.4 ± 15.5 % for solid nodules, compared with 7.6 ± 8.5 % (P < 0.01) semi-automatically. In an anthropomorphic phantom study, the sensitivity of detection is 100 % for nodules of >5 mm in diameter. Semi-automated volumetry yielded more accurate nodule volumes than manual measurements. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-012-2570-7Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 23
- Journal Issue
- 1
- Journal Page Range
- p. 139-147
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 44105834
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; AUTOMATION; CHEST; COMPUTERIZED TOMOGRAPHY; DOCUMENTATION; LUNGS; NEOPLASMS; PHANTOMS; RADIATION DOSES; RELIABILITY; SENSITIVITY; SPECIFICITY; SPHERES; VOLUME
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; MOCKUP; ORGANS; RESPIRATORY SYSTEM; STRUCTURAL MODELS; TOMOGRAPHY