The prospects for the use of X-ray computed tomography (CT) for isotherm mapping were investigated. The spatial distribution of CT changes was obtained by digital subtraction of images taken before and during heating of a 200 mm diameter cylindrical 4% agar phantom. Due to a rather unfavourable signal-to-noise ratio, the noise level in the resulting image had to be reduced by low-pass filtering using the 2-dimensional Fast Fourier Transform followed by spatial averaging. The resulting isotherms showed a resolution of 0.50C at a spatial resolution of 4.2 mm. Isotherms so obtained are presented and the possibility of using the method in clinical hyperthermia is discussed. (orig.)
Additional details
Publishing Information
- Journal Title
- Radiother. Oncol.
- Journal Volume
- 2
- Journal Issue
- 3
- Series
- Radiother. Oncol.
- Journal Page Range
- 255-260
- ISSN
- 0167-8140
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- Netherlands
- INIS RN
- 16024071
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; COMPUTERIZED TOMOGRAPHY; HYPERTHERMIA; IMAGE PROCESSING; ISOTHERMS; PHANTOMS; SPATIAL DISTRIBUTION; TEMPERATURE MEASUREMENT; X RADIATION
- Descriptors DEC
- BODY TEMPERATURE; DIAGNOSTIC TECHNIQUES; DISTRIBUTION; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MOCKUP; RADIATIONS; STRUCTURAL MODELS; TOMOGRAPHY