Published 1985 | Version v1
Report Open

Use of pressed scintillation detectors in Moessbauer spectroscopy and in nuclear technology instrumentation

Description

The method of isostatic pressing allows thin (≥0.1 mm) scintillation detectors on the base of NaI(Tl) with assigned thickness and high spectrometric parameters to be prepared. Considering an example of detection of 57Fe and 125Te Moessbauer radiation it has been shown that there is a detector optimal thickness providing maximum precision of measurement as a function of background radiation. The proposed estimation of detector optimal thickness is suitable for all cases of soft γ- or X radiation detection on the background of high-level photon radiation which occurs, for example, in γ-densitometry of radioactive solutions, X-ray spectrometry of radioactive samples etc

Availability note (English)

MF available from INIS under the Report Number.

Files

18095717.pdf

Files (332.7 kB)

Name Size Download all
md5:bb47a1efa51a338f20a278834f0e2334
332.7 kB Preview Download

System files (21.7 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Применение прессованных сцинтилляционных детекторов в мессбауэровской спектроскопии и приборах контроля ядерной технологии

Publishing Information

Imprint Pagination
10 p.
Report number
RI--178

Optional Information

Notes
5 refs.; 5 figs.