A dosimetry method for low dose rate brachytherapy by EGS5 combined with regression to reflect source strength shortage
Creators
- 1. Sapporo Medical University, Department of Medicine, Sapporo, Hokkaido (Japan)
- 2. Sapporo Medical University Hospital, Division of Radiology and Nuclear Medicine, Sapporo, Hokkaido (Japan)
- 3. Hyogo Ion Beam Medical Center, Tatsuno, Hyogo (Japan)
- 4. Teishin-kai Radiation Therapy Institute, Sapporo, Hokkaido (Japan)
Description
The post-implantation dosimetry for brachytherapy using Monte Carlo calculation by EGS5 code combined with the source strength regression was investigated with respect to its validity. In this method, the source strength for the EGS5 calculation was adjusted with the regression, so that the calculation would reproduce the dose monitored with the glass rod dosimeters (GRDs) on a water phantom. The experiments were performed, simulating the case where one of two 125I sources of Oncoseed 6711 was lacking strength by 4–48%. As a result, the calculation without regression was in agreement with the GRD measurement within 26–62%. In this case, the shortage in strength of a source was neglected. By the regression, in order to reflect the strength shortage, the agreement was improved up to 17–24%. This agreement was also comparable with accuracy of the dose calculation for single source geometry reported previously. These results suggest the validity of the dosimetry method proposed in this study. (author)
Availability note (English)
Available from http://dx.doi.org/10.1093/jrr/rrt147Additional details
Identifiers
- DOI
- 10.1093/jrr/rrt147;
Publishing Information
- Journal Title
- Journal of Radiation Research
- Journal Volume
- 55
- Journal Issue
- 3
- Journal Page Range
- p. 608-612
- ISSN
- 0449-3060
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 45090160
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S63: RADIATION, THERMAL, AND OTHER ENVIRONMENTAL POLLUTANT EFFECTS ON LIVING ORGANISMS AND BIOLOGICAL MATERIALS;
- Descriptors DEI
- BRACHYTHERAPY; DOSE RATES; DOSIMETRY; GLASS; IODINE 125; LOW LEVEL COUNTING; MONTE CARLO METHOD; RADIATION SOURCES; REGRESSION ANALYSIS; SIMULATION
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; CALCULATION METHODS; COUNTING TECHNIQUES; DAYS LIVING RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IODINE ISOTOPES; ISOTOPES; MATHEMATICS; MEDICINE; NUCLEAR MEDICINE; NUCLEI; ODD-EVEN NUCLEI; RADIOISOTOPES; RADIOLOGY; RADIOTHERAPY; STATISTICS; THERAPY
Optional Information
- Notes
- 15 refs., 3 figs., 2 tabs.