Published 1980 | Version v1
Book

Diagnostic x-ray equipment evaluation in Brazil

  • 1. Instituto de Radioprotecao e Dosimetria, Rio de Janeiro (Brazil)

Description

An outline summary is presented of the results of a survey by the Institute of Radiation Protection and Dosimetry (IRD), Atomic Energy Commission, Brazil, of 790 X-ray machines used for medical diagnosis. In a group of 120 machines in 10 hospitals, used as a sample for investigating personnel monitoring, 10% of the equipment was not functioning, due to technical failures, lack of space for proper installation or lack of staff. In a further sample of 522 machines, tests of filtration, kilovoltage, exposure time and field size gave percentages not correct as 62%, 42%, 37%, and 54%, respectively. In regard to radiation protection, 40% of the rooms for these machines did not provide adequate protection for personnel or patients, and almost all x-ray departments did not use personal dosemeters. In the field of dentistry, of 268 machines checked, only 9% were working in correct conditions of technical parameters and radiation protection. (U.K.)

Additional details

Publishing Information

Publisher
Pergamon Press.
Imprint Place
Oxford
ISBN
0 08 025912 X
Imprint Title
Radiation protection
Imprint Pagination
1278 p.
Journal Page Range
p. 169-172.

Conference

Title
5. congress of the International Radiation Protection Society on radiation protection.
Dates
Mar 1980.
Place
Jerusalem, Israel.

Optional Information

Notes
Imprint:In 2 volumes.