Published May 1988 | Version v1
Journal article

Radon in air calibration procedure: A primary method

  • 1. Argonne National Lab., 9700 S. Cass Avenue, Argonne, IL (USA)

Description

A procedure has been developed for preparing 3- to 9-iota volumes of air under natural conditions with a known concentration of /sup 222/Rn to be used for calibrating radon systems. Air is passed into a plastic bag through a standard /sup 226/Ra solution (prepared by the U.S. National Bureau of Standards) contained in an emanation flask. This plastic bag retains /sup 222/Rn with little loss into or through the bag walls. The mean ratios of the /sup 222/Rn in the air at 2 and 7 days after filling to that immediately after filling were -.992 +- 0.006 and 0.969 +- 0.008, which suggests a rate of radon loss into the bag of 0.4 +- 0.1%/day

Additional details

Publishing Information

Journal Title
Nuclear Science and Engineering
Journal Volume
99
Journal Issue
1
Series
Nucl. Sci. Eng.
Journal Page Range
82-87
ISSN
0029-5639
CODEN
NSENA