Published May 1988
| Version v1
Journal article
Radon in air calibration procedure: A primary method
Description
A procedure has been developed for preparing 3- to 9-iota volumes of air under natural conditions with a known concentration of /sup 222/Rn to be used for calibrating radon systems. Air is passed into a plastic bag through a standard /sup 226/Ra solution (prepared by the U.S. National Bureau of Standards) contained in an emanation flask. This plastic bag retains /sup 222/Rn with little loss into or through the bag walls. The mean ratios of the /sup 222/Rn in the air at 2 and 7 days after filling to that immediately after filling were -.992 +- 0.006 and 0.969 +- 0.008, which suggests a rate of radon loss into the bag of 0.4 +- 0.1%/day
Additional details
Publishing Information
- Journal Title
- Nuclear Science and Engineering
- Journal Volume
- 99
- Journal Issue
- 1
- Series
- Nucl. Sci. Eng.
- Journal Page Range
- 82-87
- ISSN
- 0029-5639
- CODEN
- NSENA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 20015342
- Subject category
- S54: ENVIRONMENTAL SCIENCES;
- Descriptors DEI
- AIR POLLUTION; CALIBRATION; ENVIRONMENTAL TRANSPORT; FEASIBILITY STUDIES; MONITORING; RADIUM 226; RADON 222; US NBS
- Descriptors DEC
- ALPHA DECAY RADIOISOTOPES; CARBON 14 DECAY RADIOISOTOPES; DAYS LIVING RADIOISOTOPES; EVEN-EVEN NUCLEI; HEAVY ION DECAY RADIOISOTOPES; HEAVY NUCLEI; ISOTOPES; MASS TRANSFER; NATIONAL ORGANIZATIONS; NUCLEI; POLLUTION; RADIOISOTOPES; RADIUM ISOTOPES; RADON ISOTOPES; US ORGANIZATIONS; YEARS LIVING RADIOISOTOPES