Published October 1982 | Version v1
Report Open

Mass-resolving charge-exchange system on PDX

Description

The PDX charge-exchange system is comprised of four, ten channel, mass-resolved, charge-exchange analyzers. Each analyzer is constructed with parallel electric and magnetic fields and is calibrated over an energy range of 0.5 to 40 keV. The mass rejection between hydrogen and deuterium has been measured as better than 1000 to 1. For ohmic heated discharges the system can provide single shot radial ion temperature profiles (4 point) with 1 msec time resolution. For neutral beam heated discharges complete radial and temporal profiles can be obtained in 2 to 4 shots. The system is also equipped with a vertically aimed diagnostic neutral beam to allow local ion-energy distribution measurements. This report describes the analyzer system and its calibration, and presents results from ohmic and neutral beam heated discharges

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A04/MF A01 as DE83002336.

Files

14744103.pdf

Files (749.5 kB)

Name Size Download all
md5:93f3d2c594f6232c6f1aa19ad1e95ac3
749.5 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
59 p.
Report number
PPPL--1940

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14744103
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
Descriptors DEI
ATOMIC BEAMS; CALIBRATION; CHARGE EXCHANGE; ELECTRIC FIELDS; ION TEMPERATURE; MAGNETIC FIELDS; PDX DEVICES; PLASMA DIAGNOSTICS
Descriptors DEC
BEAMS