An integrated Monte Carlo dosimetric verification system for radiotherapy treatment planning
Creators
- 1. Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Osaka 565-0871 (Japan)
- 2. Department of Radiation Oncology and Image-Applied Therapy, Kyoto University Graduate School of Medicine, Sakyo, Kyoto 606-8507 (Japan)
- 3. Clinical Radiology Service Division, Kyoto University Hospital, Sakyo, Kyoto 606-8507 (Japan)
Description
An integrated Monte Carlo (MC) dose calculation system, MCRTV (Monte Carlo for radiotherapy treatment plan verification), has been developed for clinical treatment plan verification, especially for routine quality assurance (QA) of intensity-modulated radiotherapy (IMRT) plans. The MCRTV system consists of the EGS4/PRESTA MC codes originally written for particle transport through the accelerator, the multileaf collimator (MLC), and the patient/phantom, which run on a 28-CPU Linux cluster, and the associated software developed for the clinical implementation. MCRTV has an interface with a commercial treatment planning system (TPS) (Eclipse, Varian Medical Systems, Palo Alto, CA, USA) and reads the information needed for MC computation transferred in DICOM-RT format. The key features of MCRTV have been presented in detail in this paper. The phase-space data of our 15 MV photon beam from a Varian Clinac 2300C/D have been developed and several benchmarks have been performed under homogeneous and several inhomogeneous conditions (including water, aluminium, lung and bone media). The MC results agreed with the ionization chamber measurements to within 1% and 2% for homogeneous and inhomogeneous conditions, respectively. The MC calculation for a clinical prostate IMRT treatment plan validated the implementation of the beams and the patient/phantom configuration in MCRTV
Additional details
Identifiers
- DOI
- 10.1088/0031-9155/52/7/014;
- PII
- S0031-9155(07)32455-X;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 52
- Journal Issue
- 7
- Journal Page Range
- p. 1991-2008
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38072569
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCELERATORS; IONIZATION CHAMBERS; LUNGS; MONTE CARLO METHOD; PATIENTS; PHANTOMS; PHASE SPACE; PHOTON BEAMS; PLANNING; PROSTATE; QUALITY ASSURANCE; RADIOTHERAPY; SKELETON; VERIFICATION
- Descriptors DEC
- BEAMS; BODY; CALCULATION METHODS; GLANDS; MALE GENITALS; MATHEMATICAL SPACE; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIATION DETECTORS; RADIOLOGY; RESPIRATORY SYSTEM; SPACE; STRUCTURAL MODELS; THERAPY