Mammography practice in Serbia: Evaluation and optimisation of image quality and the technical aspects of the mammographic imaging chain
Creators
- 1. Radiation Protection Laboratory, Radiation and Environmental Protection Dept., Vinca Inst. of Nuclear Sciences, M. P. Alasa 12-14, Vinca, Belgrade (Serbia)
Description
The purpose of this work was to assess mammography practice in Serbia and its appropriateness for both diagnostic service and potential screening by implementing quality control (QC) protocol in three large teaching hospitals. Corrective actions were suggested, accordingly. In addition to technical aspects of QC, image quality was assessed using image grading before and after the introduction of corrective measures. The survey demonstrated considerable variations in technical parameters that affect image quality and patients doses. Average glandular doses ranged from 1.8 to 2.8 mGy, while reference optical density (OD) ranged from 1.0 to 2.6. Image grading resulted in a very high percentage of images with poor quality (12-70% for cranio-caudal projection and 8-66% for medio-lateral oblique projection). Main problems were associated with film processing, viewing conditions and OD control. Following introduction of corrective measures, the image grading results were improved in some hospitals, so the percentage of images without any remarks has been increased. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncq096Additional details
Identifiers
- DOI
- 10.1093/rpd/ncq096;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 139
- Journal Issue
- 1-3
- Journal Page Range
- p. 293-297
- ISSN
- 0144-8420
Conference
- Title
- 3. Malmoe Conference on Medical Imaging - Optimisation in X-ray and molecular imaging
- Dates
- 25-27 Jun 2009
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43102244
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DOSES; EVALUATION; FEMALES; FILMS; HUMAN POPULATIONS; IMAGES; MAMMARY GLANDS; NEOPLASMS; PATIENTS; QUALITY ASSURANCE; QUALITY CONTROL; SCREENING; SENSITIVITY; SPECIFICITY
- Descriptors DEC
- BODY; CONTROL; DIAGNOSTIC TECHNIQUES; DISEASES; GLANDS; MEDICINE; NUCLEAR MEDICINE; ORGANS; POPULATIONS; RADIOLOGY
Optional Information
- Notes
- 10 refs