Published June 1985
| Version v1
Journal article
Caesium ion-selective electrode based on 15-crown-5-phospho-tungstic acid precipitates
Description
A caesium ion-selective PVC membrane electrode based on 15-crown-5-phosphotungstic acid (PW) precipitates as the neutral carrier was prepared. The electrode gave a linear response with a Nernstian slope of 60 mv decade-1 within the concentration range 10-1 to 10-4 M CsNO3. The selectivity and response time of the electrode were investigated and it was found that the electrode exhibited good selectivity for caesium in comparison with alkali, alkaline earth and some transition metal ions. The effects of pH, addition of ethanol and anions present in the test solutions were also studied. (author)
Additional details
Publishing Information
- Journal Title
- Analyst
- Journal Volume
- 110
- Journal Issue
- 6
- Series
- Analyst.
- Journal Page Range
- 635-638
- ISSN
- 0003-2654
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16070205
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- AQUEOUS SOLUTIONS; CESIUM; ELECTRODES; ELECTROLYSIS; ION SELECTIVE ELECTRODE ANALYS; MEMBRANES; PH VALUE; POLYETHYLENE GLYCOLS; PVC; QUANTITATIVE CHEMICAL ANALYSIS; TETRAHYDROFURAN; TITRATION; TUNGSTOPHOSPHORIC ACID
- Descriptors DEC
- ALKALI METALS; CHEMICAL ANALYSIS; DISPERSIONS; ELEMENTS; FURANS; HETEROCYCLIC COMPOUNDS; HOMOGENEOUS MIXTURES; HYDROGEN COMPOUNDS; INORGANIC ACIDS; METALS; MIXTURES; ORGANIC CHLORINE COMPOUNDS; ORGANIC COMPOUNDS; ORGANIC HALOGEN COMPOUNDS; ORGANIC OXYGEN COMPOUNDS; ORGANIC POLYMERS; OXYGEN COMPOUNDS; PHOSPHORUS COMPOUNDS; POLYMERS; POLYVINYLS; SOLUTIONS; TRANSITION ELEMENT COMPOUNDS; TUNGSTEN COMPOUNDS