The influence of the energy distribution of workplace fields on neutron personal dosemeter reading
Description
Variations in the energy dependence of response of neutron personal dosemeters cause systematic errors in the readings obtained in workplace fields. The magnitude of these errors has been determined theoretically by folding measured and calculated workplace energy distributions with dosemeter response functions, to determine the response of a given personal dosemeter in that field. These results have been analysed with consideration of the dosemeter response to various calibration spectra, and with reference to different workplaces. The dosemeters in the study are discussed in terms of the workplaces for which they can be suitably calibrated. Deficiencies in the published neutron energy distributions are identified
Additional details
Identifiers
- PII
- S0168900201014796;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 476
- Journal Issue
- 1-2
- Journal Page Range
- p. 405-410
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34001090
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CALIBRATION; ENERGY SPECTRA; NEUTRON DOSIMETRY; PERSONNEL DOSIMETRY; RADIATION PROTECTION; RESPONSE FUNCTIONS
- Descriptors DEC
- DOSIMETRY; FUNCTIONS; SPECTRA
Optional Information
- Copyright
- Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.