Published June 1979
| Version v1
Journal article
Measurement of the energy of high-energy electron beam with rapid-processed film
Creators
- 1. Seoul National University College of Medicine, Seoul (Korea, Republic of)
Description
Ready packed, rapid- processable film exposed within a polystyrene phantom affords a rapid, simple and accurate method of determining electron beam energy from the range-energy relationship. For accelerators where there may be instability during the early portion of exposure, the use of the slow RP/V film is preferable because most the film exposure, the use of the produced under stable operating conditions. A particular advantage of the method is the amount of time required is in the order of typical treatment time.
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Radiological Society
- Journal Volume
- 15
- Journal Issue
- 1
- Series
- 6 refs, 3 figs, 1 tab
- Journal Page Range
- p. 292-294
- ISSN
- 1738-2637
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 41021069
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCELERATORS; ELECTRON BEAMS; ENERGY; PHANTOMS
- Descriptors DEC
- BEAMS; LEPTON BEAMS; MOCKUP; PARTICLE BEAMS; STRUCTURAL MODELS