Published March 1979 | Version v1
Journal article

The fixed horizontal neutron therapy beam at Edinburgh: dosimetry and radiation protection

  • 1. Western General Hospital, Edinburgh (UK)

Description

A compact cyclotron producing 15 MeV deutrons has been installed at the Western General Hospital, Edinburgh, to extend the Medical Research Council's clinical trails of fast neutrons. Two treatment rooms are available one of which has an isocentric unit and the other a fixed horizontal beam which is the subject of this paper. A radiation protection survey has demonstrated safe levels of radiation throughout the building although there is some activation of the fixed horizontal beam cone resulting in doses to radiographers of 10 mrem per week. The calibration of neutron dose is discussed and the measurements of dose distribution dexribed. Isodose and depth-dose data for both the neutron and photon components of the field are presented. (author)

Additional details

Publishing Information

Journal Title
Br. J. Radiol.
Journal Volume
52
Journal Issue
615
Series
Br. J. Radiol.
Journal Page Range
197-208
ISSN
0007-1285