Published 1996 | Version v1
Book

A low DC drift read-out system for a large mass bolometric detector

  • 1. Istituto Nazionale di Fisica Nucleare, Milano (Italy)

Description

We present a read-out solution for a bolometric detector of large mass. It consist of a voltage-sensitive preamplifier having very low series and parallel input noise, thanks to the use of a pair of selected silicon FET at the input. Very good electrical characteristics were obtained. DC power supply rejection ratio, thermal drift of input offset voltage and input leakage current are minimized by using novel circuit solutions. Parallel noise behavior was also investigated using a new measurement method capable to be sensitive to shot noise generated by very small currents (tens of fA)

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (United States)
Imprint Title
1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
Imprint Pagination
2138 p.
Journal Page Range
p. 504-508.

Conference

Title
Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1996.
Place
Anaheim, CA (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
28068899
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S42: ENGINEERING;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BOLOMETERS; DESIGN; FIELD EFFECT TRANSISTORS; LEAKAGE CURRENT; PREAMPLIFIERS; READOUT SYSTEMS
Descriptors DEC
AMPLIFIERS; CURRENTS; ELECTRIC CURRENTS; ELECTRONIC EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS; SEMICONDUCTOR DEVICES; TRANSISTORS

Optional Information

Secondary number(s)
CONF-961123--.