A new method for the correction of gamma camera nonuniformity due to spatial distortion
Creators
- 1. Department of Radiology C-278, University of Colorado Health Sciences Center, Denver, CO 80262 (United States)
- 2. Cardiology Associates, PC, Denver, CO 80110 (United States)
Description
A methodology for correcting scintillation camera nonuniformity resulting from spatial linearity distortion is described. The method simultaneously corrects for both variations in count density and distortions in spatial linearity without altering the intrinsic sensitivity or resolution of the imaging system. This approach to linearity correction requires only a single flood image from which the spatial shift vectors are derived. The algorithm has been implemented on a PC-based, EISA bus microcomputer, and reduces the measurement of integral uniformity for 201Tl from 25% to 5%. Combined with regional correction for photopeak pulse-height shifts (i.e. energy correction), the algorithm represents a novel technique for implementing uniformity correction completely within software. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 41
- Journal Issue
- 10
- Journal Page Range
- p. 2179-2188
- ISSN
- 1361-6560
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 31025306
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPUTER CODES; DIAGNOSTIC USES; GAMMA CAMERAS; SCINTISCANNING; SENSITIVITY; SPATIAL RESOLUTION
- Descriptors DEC
- CAMERAS; COUNTING TECHNIQUES; DIAGNOSTIC TECHNIQUES; RADIOISOTOPE SCANNING; RESOLUTION; USES
Optional Information
- Notes
- Refs