Optimization of the obtaining method of CaSO4: Dy + Ptfe dosimeters

Description

This work contain the obtained results of studying the response when irradiating to different dose of X-rays, dosimeters of CaSO4: Dy + Ptfe using different lapses of time in their preparation (a lot of selected dosimeters of an elaboration process of 3 days and another of an elaboration process of 2 hours). For the elaboration of the powdered material, the evaporation method was used; the irradiation were carried out in a lineal accelerator Elekta Synergy property of the National Medical Center, 20 de November. The similarities and differences are shown among the two dosimeters lots together with an analysis of the shine curves and of calibration selecting those that presented a better behaviour and a more rea liable response. (Author)

Files

44026231.pdf

Files (410.6 kB)

Name Size Download all
md5:a1b4308bb266e0b634efb730eb04b050
410.6 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Optimizacion del metodo de obtencion de dosimetros de CaSO

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico D. F. (Mexico)
ISBN
978-607-00-6167-7
Imprint Pagination
5 p.
Report number
INIS-MX--2672

Conference

Title
13. International Symposium; 23. National Congress on Solid State Dosimetry
Original Conference Title
13. Conferencia Internacional
Dates
15-19 Oct 2012
Place