Published July 2015 | Version v1
Journal article

Developments in time-resolved high pressure x-ray diffraction using rapid compression and decompression

  • 1. High Pressure Collaborative Access Team, Geophysical Laboratory, Carnegie Institution of Washington, Argonne, Illinois 60439 (United States)

Description

Complementary advances in high pressure research apparatus and techniques make it possible to carry out time-resolved high pressure research using what would customarily be considered static high pressure apparatus. This work specifically explores time-resolved high pressure x-ray diffraction with rapid compression and/or decompression of a sample in a diamond anvil cell. Key aspects of the synchrotron beamline and ancillary equipment are presented, including source considerations, rapid (de)compression apparatus, high frequency imaging detectors, and software suitable for processing large volumes of data. A number of examples are presented, including fast equation of state measurements, compression rate dependent synthesis of metastable states in silicon and germanium, and ultrahigh compression rates using a piezoelectric driven diamond anvil cell

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
86
Journal Issue
7
Journal Page Range
p. 072208-072208.9
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2015 AIP Publishing LLC