Published April 1, 2008
| Version v1
Journal article
Automated measurement of the refractive index of fluids
Creators
Description
A method for automating refractive-index measurements of fluids has been developed. An encoded rotation stage and position-sensitive detector enable automated reading of angles typically acquired by visual means. Two tunable lasers are used to obtain index measurements at ten discrete wavelengths across the visible spectrum. This method has been implemented on a Hilger-Chance refractometer from which the bulk refractive-index values for various transparent fluids have been measured. An index measurement accuracy of better than one part in the fourth decimal place for distilled water and a few parts in the fourth decimal place for higher index fluids is obtained
Additional details
Identifiers
- DOI
- 10.1364/AO.47.001504;
Publishing Information
- Journal Title
- Applied Optics
- Journal Volume
- 47
- Journal Issue
- 10
- Journal Page Range
- p. 1504-1509
- ISSN
- 0003-6935
- CODEN
- APOPAI
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 40003230
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Descriptors DEI
- FLUIDS; LASERS; POSITION SENSITIVE DETECTORS; REFRACTIVE INDEX; ROTATION; VISIBLE SPECTRA; WAVELENGTHS
- Descriptors DEC
- MEASURING INSTRUMENTS; MOTION; OPTICAL PROPERTIES; PHYSICAL PROPERTIES; RADIATION DETECTORS; SPECTRA
Optional Information
- Notes
- (c) 2008 Optical Society of America