Novel transmission curved crystal spectrometer used on hard X-ray diagnoses
Creators
- 1. China Academy of Engineering Physics, Mianyang (China). Research Center of Laser Fusion
Description
A novel transmission curved crystal spectrometer is described for hard X-ray measurements in the energy range of 10-60 keV. The spectrometer is based on a cylindrical curved quartz crystal. The curvature radius is 112 mm and the spacing of the crystal lattice is 0.4256 nm. The experiment utilizes a Mo X-ray tube as the light source and the spectrometer is set perpendicular to the axis about 700 mm distance. Two characteristic spectra are symmetrically recorded by image plate clearly and the experiment is well consistent with theoretical analysis. The error analysis shows that the aim accuracy is the main factor. (authors)
Additional details
Publishing Information
- Journal Title
- High Power Laser and Particle Beams
- Journal Volume
- 20
- Journal Issue
- 9
- Journal Page Range
- p. 1491-1494
- ISSN
- 1001-4322
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 40104846
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- ACCURACY; CRYSTAL LATTICES; CRYSTALS; CYLINDRICAL CONFIGURATION; DISTANCE; ERRORS; HARD X RADIATION; IMAGES; KEV RANGE; LIGHT SOURCES; MOLYBDENUM; PLATES; QUARTZ; SPECTROMETERS; TARGETS; TRANSMISSION; X-RAY DETECTION; X-RAY SPECTRA; X-RAY TUBES
- Descriptors DEC
- CONFIGURATION; CRYSTAL STRUCTURE; DETECTION; ELECTROMAGNETIC RADIATION; ELECTRON TUBES; ELEMENTS; ENERGY RANGE; EQUIPMENT; IONIZING RADIATIONS; MEASURING INSTRUMENTS; METALS; MINERALS; OXIDE MINERALS; RADIATION DETECTION; RADIATION SOURCES; RADIATIONS; REFRACTORY METALS; SPECTRA; TRANSITION ELEMENTS; X RADIATION; X-RAY EQUIPMENT
Optional Information
- Notes
- 3 figs., 3 tabs., 6 refs.