Published February 1987 | Version v1
Journal article

A measurement technique for trace radioactivity distributions in the human body by using gamma-ray spectra

  • 1. Energy Research Lab., Hitachi Ltd., 1168 Moriyama-cho, Hitachi-shi, Ibaraki-ken, 316

Description

A new technique is described to increase spatial resolution in the distribution measurement on amount of trace radioactivity in the human body by the response function method. The distribution is obtained by analysing the difference in gamma-ray spectra measured around the body, which contains information for not only unscattered gamma-rays, but also scattered ones within the body. Basic experiments involving the measurement of two-dimensional radioactivity distribution in a homogeneous acrylate block (40cmW x 20cmH x 50cmL) containing two /sup 137/Cs point sources (0.8 and 1.0 μCi) were made using gamma-ray counts measured by twenty 2'' phi x 2'' NaI(Tl) scintillators arranged on a circle, and the response function calculated by the Monte Carlo technique. The distribution on a 10 x 5 mesh division was successfully analysed with a relative error of less than +-20% for a measuring time of 10 min. The mesh size was 4cm x 4cm, which was about one-half that possible when analysing only unscattered rays

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-34
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
611-615
ISSN
0018-9499
CODEN
IETNA

Conference

Title
Nuclear science and nuclear power systems symposium.
Dates
29-31 Oct 1986.
Place
Washington, DC (USA).

Optional Information

Secondary number(s)
CONF-861007--.