Published May 1984 | Version v1
Journal article

Absolute left ventricular volume by an iterative build-up factor analysis of gated radionuclide images

Description

A method for determining absolute left ventricular (LV) volumes from radionuclide gated blood-pool (GBP) images was validated in 34 patients. The technique is nongeometric, corrects for tissue attenuation, and uses an experimentally determined set of build-up factors to account for the effects of scatter. Only four parameters are needed to determine LV volumes: the LV count rates from a left anterior oblique (LAO) and a right posterior oblique (RPO) image (1800 opposed to the LAO), a venous blood sample, and a patient thickness measurement. A computer algorithm is used to reach an iterative solution to two simultaneous equations that yield LV volumes. Phantom studies showed less than 4% error for volume determinations at all investigated depths. For the patients studied the correlation between volumes obtained by GBP contrast ventriculography was 0.97 for diastole and 0.96 for systole

Additional details

Publishing Information

Journal Title
Radiology
Journal Volume
151
Journal Issue
2
Series
Radiology.
Journal Page Range
477-481