Automatic measurement of oxygen isotope ratio of water samples
- 1. Okayama Univ., Misasa, Tottori (Japan). Inst. for Thermal Spring Research
Description
The automatic sample preparation system for oxygen isotope analysis of natural water samples was constructed. The system is essentially a modification of that originally designed by W. Dansgaard in the University of Copenhagen. Sixty water samples of 5 ml each are automatically equilibrated with CO2 of 30 ml NTP each within 4.5 hours. The equilibrated CO2 gases are successively measured for thier oxygen isotopic ratios by an automatic mass-spectrometer, VG903. The time required for the measurement of the 60 CO2 samples is about 15 hours; thus approximately 20 hours are required to complete the isotopic measurements of 60 water samples. The accuracy of the oxygen isotopic analyses is about 0.1 %. The result of oxygen isotopic analyses by the automatic preparation system agrees with that of the conventional method within about +0.1 %. The results indicate that the automatic preparation system is applicable for the oxygen isotope analysis of natural water samples. (Kubozono, M.)
Additional details
Publishing Information
- Journal Title
- Okayama Daigaku Onsen Kenkyusho Hokoku
- Journal Issue
- no.56
- Series
- Okayama Daigaku Onsen Kenkyusho Hokoku.
- ISSN
- 0369-7142
- CODEN
- ODOKA
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 18017436
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- ACCURACY; AUTOMATION; FLOWSHEETS; GEOLOGY; GROUND WATER; ISOTOPE RATIO; MASS SPECTROMETERS; MEASURING METHODS; OXYGEN 18; PERFORMANCE; SAMPLE CHANGERS; SAMPLE PREPARATION
- Descriptors DEC
- DIAGRAMS; EVEN-EVEN NUCLEI; HYDROGEN COMPOUNDS; INFORMATION; ISOTOPES; LIGHT NUCLEI; MEASURING INSTRUMENTS; NUCLEI; OXYGEN COMPOUNDS; OXYGEN ISOTOPES; POLAR SOLVENTS; SOLVENTS; SPECTROMETERS; STABLE ISOTOPES; WATER