Published May 1998 | Version v1
Journal article

Peripheral Insertion of a Central Venous Access Device Under Fluoroscopic Guidance Using a Peripherally Accessed System (PAS) Port in the Forearm

  • 1. Department of Radiology, Kochi Municipal Central Hospital, 2-7-33 Sakurai-cho, Kochi-city 780 (Japan)
  • 2. Department of Gastroenterology, Kochi Municipal Central Hospital, 2-7-33 Sakurai-cho, Kochi-city 780 (Japan)
  • 3. Department of Surgery, Kochi Municipal Central Hospital, 2-7-33 Sakurai-cho, Kochi-city 780 (Japan)
  • 4. Department of Radiology, Hyogo Medical Center for Adults, 13-70 Kitaouji-cho, Akashi-city, Hyogo 673 (Japan)

Description

Purpose: We describe the technique, efficacy, and complications of fluoroscopy-guided implantation of a central venous access device using a peripherally accessed system (PAS) port via the forearm. Methods: Beginning in July 1994, 105 central venous access devices were implanted in 104 patients for the long-term infusion of antibiotics or antineoplasmic agents, blood products, or parenteral nutrition. The devices was inserted under fluoroscopic guidance with real-time venography from a peripheral route. Results: All ports were successfully implanted. There were no procedure-related complications. No thrombosis or local infection was observed; however, in six patients catheter-related phlebitis occurred. Conclusion: Fluoroscopy-guided implantation of a central venous access device using a PAS port via the forearm is safe and efficacious, and injection of contrast medium through a peripheral IV catheter before introduction of the catheter helps to avoid catheter-related phlebitis

Additional details

Identifiers

Publishing Information

Journal Title
Cardiovascular and Interventional Radiology
Journal Volume
21
Journal Issue
3
Journal Page Range
p. 230-233
ISSN
0174-1551
CODEN
CAIRDG

Optional Information

Copyright
Copyright (c) 1996 Springer-Verlag New York Inc.