Digital switch boards for IV and CV measurements of large-array Low Gain Avalanche Detectors
Creators
- 1. State Key Laboratory of Particle Detection and Electronics, University of Science and Technology of China, Hefei 230026 (China)
- 2. Department of Modern Physics, University of Science and Technology of China, Hefei 230026 (China)
- 3. SCIPP, University of California Santa Cruz, CA 95064 (United States)
Description
The High-Granularity Timing Detector (HGTD), with a time resolution of approximately 30 ps, is proposed for the ATLAS Phase II upgrade to address the challenge of greatly increased pile-up interactions. The HGTD is based on the technology of Low Gain Avalanche Detectors (LGADs). The current–voltage (IV) and capacitance–voltage (CV) measurements are important methods to study the main characteristics that can be exploited for quality controlling LGADs during production. A single probe can be controlled via computer to switch channels when using an automatic probe station, but this leaves the remaining, adjacent channels floating and results in inaccurate measurements due to the punch-through effect. We designed the digital switch boards for the IV and CV measurements of 5 5 and 15 15 LGADs arrays. With the digital switch board and probe cards, channels on a large-array LGADs can be scanned automatically during the measurement, which greatly improves the measurement efficiency. The results show that the leakage currents and noise of all channels induced by the two digital switch boards are lower than 10 pA. We show the IV and CV curves of all channels on two LGADs from Hamamatsu Photonics K. K. (HPK) with our digital switch boards. Compared with the results measured by HPK with an automatic probe station, the IV and CV curves that we measured have better uniformity and accuracy, demonstrating that the quality control of the large-array LGADs can be conducted well with our digital switch board.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2021.165400Additional details
Identifiers
- DOI
- 10.1016/j.nima.2021.165400;
- PII
- S0168900221003843;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 1005
- Journal Page Range
- vp.
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54086472
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Descriptors DEI
- CAPACITANCE; COMPUTERS; DESIGN; ELECTRIC POTENTIAL; LEAKAGE CURRENT; QUALITY CONTROL; TIME RESOLUTION
- Descriptors DEC
- CONTROL; CURRENTS; ELECTRIC CURRENTS; ELECTRICAL PROPERTIES; PHYSICAL PROPERTIES; RESOLUTION; TIMING PROPERTIES
Optional Information
- Copyright
- Copyright (c) 2021 Elsevier B.V. All rights reserved.