Energy calibration for the INDRA multidetector using recoil protons from 12C+1H scattering

Description

An efficient method of energy scale calibration for the CsI(Tl) modules of the INDRA multidetector (rings 6-12) using elastic and inelastic 12C+1H scattering at E(12C)=30 MeV per nucleon is presented. Background-free spectra for the binary channels are generated by requiring the coincident detection of the light and heavy ejectiles. The gain parameter of the calibration curve is obtained by fitting the proton total charge spectra to the spectra predicted with Monte-Carlo simulations using tabulated cross section data. The method has been applied in multifragmentation experiments with INDRA at GSI

Additional details

Identifiers

PII
S0168900203006235;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
501
Journal Issue
2-3
Journal Page Range
p. 367-374
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.