Published May 11, 1998 | Version v1
Journal article

Quantitative consideration of collimator parameters in radio-isotope imaging and its application to collimator optimization

Description

A Monte Carlo aided method of calculating the collimator performance parameters is presented and used to quantify the effect of collimator dimensions and photon energy in emission planar and tomographic imaging. The importance of selecting the optimum collimator length and bore diameter is shown and the significance of collimator edge penetration at higher photon energies is demonstrated. Using image quality parameters a practical method of designing a suitable parallel-hole collimator is developed and the trade-offs that must be considered for a successful design are discussed

Additional details

Identifiers

PII
S0969804397100987;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
49
Journal Issue
8
Journal Page Range
p. 939-954
ISSN
0969-8043
CODEN
ARISEF

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34013521
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
COLLIMATORS; DESIGN; MONTE CARLO METHOD; OPTIMIZATION; RADIOTHERAPY; TOMOGRAPHY
Descriptors DEC
CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 1998 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.