Calibration of high order integral diffraction coefficient for X-ray PET flat crystal on the automatic X-ray diffractometer
- 1. Research Center of Laser Fusion, CAEP, Mianyang (China)
Description
On the automatic X-ray diffractometer (XRD), based on the stability and precision control of θ and 2θ goniometer, special plane crystal holder was made. Bremsstrahlung and Cu Kβ line are attenuated for 5 orders by 40 μm-thick pure Nickel filter in 15 kilo-voltage and 20 mA supply to X-ray Cu tube, X-ray source of Cu target is to be Cu Kα monochromatic source, transmission coefficient of Nickel filter is the criterion of Cu Kα monochromatic source. For X-ray Pentaerythritol (002) crystal, integral diffraction coefficient of 1-3 order diffraction on Cu Kα energy were calibrated. The results show that the 2nd order and the 3rd order is only 14.36% and 4.07%, the Cu Kα1 intensity ratio of 2nd order and the 3rd order is only 7.7% and 1.3%, comparing the 1st order diffraction. This kind of calibration is efficient and convenient on XRD in common laboratory. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 33
- Journal Issue
- 12
- Journal Page Range
- p. 1538-1542
- ISSN
- 0258-0934
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 49064801
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BREMSSTRAHLUNG; CALIBRATION; COMPARATIVE EVALUATIONS; CRYSTALS; ELECTRIC POTENTIAL; FILTERS; MONOCHROMATIC RADIATION; POSITRON COMPUTED TOMOGRAPHY; X-RAY DIFFRACTION; X-RAY DIFFRACTOMETERS; X-RAY SOURCES
- Descriptors DEC
- COHERENT SCATTERING; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DIFFRACTION; DIFFRACTOMETERS; ELECTROMAGNETIC RADIATION; EMISSION COMPUTED TOMOGRAPHY; EVALUATION; MEASURING INSTRUMENTS; RADIATION SOURCES; RADIATIONS; SCATTERING; TOMOGRAPHY
Optional Information
- Notes
- 7 figs., 3 tabs., 6 refs.