Published December 2013 | Version v1
Journal article

Calibration of high order integral diffraction coefficient for X-ray PET flat crystal on the automatic X-ray diffractometer

  • 1. Research Center of Laser Fusion, CAEP, Mianyang (China)

Description

On the automatic X-ray diffractometer (XRD), based on the stability and precision control of θ and 2θ goniometer, special plane crystal holder was made. Bremsstrahlung and Cu Kβ line are attenuated for 5 orders by 40 μm-thick pure Nickel filter in 15 kilo-voltage and 20 mA supply to X-ray Cu tube, X-ray source of Cu target is to be Cu Kα monochromatic source, transmission coefficient of Nickel filter is the criterion of Cu Kα monochromatic source. For X-ray Pentaerythritol (002) crystal, integral diffraction coefficient of 1-3 order diffraction on Cu Kα energy were calibrated. The results show that the 2nd order and the 3rd order is only 14.36% and 4.07%, the Cu Kα1 intensity ratio of 2nd order and the 3rd order is only 7.7% and 1.3%, comparing the 1st order diffraction. This kind of calibration is efficient and convenient on XRD in common laboratory. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
33
Journal Issue
12
Journal Page Range
p. 1538-1542
ISSN
0258-0934

Optional Information

Notes
7 figs., 3 tabs., 6 refs.