Published March 1986 | Version v1
Journal article

Absorbed dose calorimetry: Bridge nulling and voltage analysis techniques

  • 1. Department of Radiation Medicine, Massachusetts General Hospital Cancer Center, Harvard Medical School, Boston, Massachusetts 02114

Description

A theoretical analysis is made of the conventional calorimeter approach utilizing a two-thermistor Wheatstone bridge: an approach which requires balancing the bridge by adjusting variable resistances. However, with the current availability of computers and digital precision measuring instruments, the conventional calorimeter approach can now be readily replaced by an automated, nonmechanical, electrical systems approach called differential calorimetry. The analytical theory outlining such an approach is also presented

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
13
Journal Issue
2
Series
Med. Phys.
Journal Page Range
242-247
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17051928
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALORIMETRIC DOSEMETERS; CALORIMETRY; RADIOTHERAPY; SPECIFICATIONS; TEMPERATURE MEASUREMENT
Descriptors DEC
DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; THERAPY