Published March 1986
| Version v1
Journal article
Absorbed dose calorimetry: Bridge nulling and voltage analysis techniques
Creators
- 1. Department of Radiation Medicine, Massachusetts General Hospital Cancer Center, Harvard Medical School, Boston, Massachusetts 02114
Description
A theoretical analysis is made of the conventional calorimeter approach utilizing a two-thermistor Wheatstone bridge: an approach which requires balancing the bridge by adjusting variable resistances. However, with the current availability of computers and digital precision measuring instruments, the conventional calorimeter approach can now be readily replaced by an automated, nonmechanical, electrical systems approach called differential calorimetry. The analytical theory outlining such an approach is also presented
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 13
- Journal Issue
- 2
- Series
- Med. Phys.
- Journal Page Range
- 242-247
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17051928
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALORIMETRIC DOSEMETERS; CALORIMETRY; RADIOTHERAPY; SPECIFICATIONS; TEMPERATURE MEASUREMENT
- Descriptors DEC
- DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; THERAPY