Intensity-modulated radiation therapy to bilateral lower limb extremities concurrently: a planning case study

  • 1. Radiation Oncology, Epworth HealthCare, Victoria (Australia)

Description

Non-melanomatous skin cancers represent 80% of all newly diagnosed cancers in Australia with basal cell carcinoma (BCC) and squamous cell carcinoma (SCC) being the most common. A previously healthy 71-year-old woman presented with widespread and tender superficial skin cancers on the lower bilateral limbs. External beam radiation therapy through the use of intensity-modulated radiation therapy (IMRT) was employed as the treatment modality of choice as this technique provides conformal dose distribution to a three-dimensional treatment volume while reducing toxicity to surrounding tissues. The patient was prescribed a dose of 60 Gy to the planning target volume (PTV) with 1.0 cm bolus over the ventral surface of each limb. The beam arrangement consisted of six treatment fields that avoided entry and exit through the contralateral limb. The treatment plans met the International Commission on Radiation Units and Measurements (ICRU) guidelines and produced highly conformal dosimetric results. Skin toxicity was measured against the National Cancer Institute: Common Terminology Criteria for Adverse Events (NCI: CTCAE) version 3. A well-tolerated treatment was delivered with excellent results given the initial extent of the disease. This case study has demonstrated the feasibility and effectiveness of IMRT for skin cancers as an alternative to surgery and traditional superficial radiation therapy, utilising a complex PTV of the extremities for patients with similar presentations

Availability note (English)

Available from http://dx.doi.org/10.1002/jmrs.57; Available from http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4175847

Additional details

Publishing Information

Journal Title
Journal of Medical Radiation Sciences (Print)
Journal Volume
61
Journal Issue
3
Journal Page Range
p. 210-215
ISSN
2051-3895

Optional Information

Copyright
Copyright (c) 2014 The Authors. Journal of Medical Radiation Sciences published by Wiley Publishing Asia Pty Ltd on behalf of Australian Institute of Radiography and New Zealand Institute of Medical Radiation Technology
Notes
PMCID: PMC4175847; PMID: 26229657; OAI: oai:pubmedcentral.nih.gov:4175847