Survey of computed tomography technique and radiation dose in Sudanese hospitals
- 1. Radiation Safety Institute, Sudan Atomic Energy Commission, P.O. Box 3001, Khartoum (Sudan)
- 2. Department of Physics, Faculty of Science, University of Khartoum, P.O. Box 321, Khartoum (Sudan)
Description
The purpose of this study was to survey technique and radiation absorbed dose in CT examinations of adult in Sudan and to compare the results with the reference dose levels. Questionnaire forms were completed in nine hospitals and a sample of 445 CT examinations in patients. Information on patient, procedure, scanner, and technique for common CT examinations were collected. For each facility, the radiation absorbed dose was measured on CT dose phantom measuring 16 cm (head) and 32 cm (body) in diameter and was used to calculate the normalized CT air kerma index. Volume CT air kerma index (CVOL), CT air kerma-length product (PKL,CT) values were calculated using the measured normalized CT air kerma index and questionnaire information. The effective dose, E estimates was determined by using PKL,CT measurements and appropriate normalized coefficients. Assuming the sample to offer a fair representative picture of CT practice patterns in Sudan, the mean CVOL and PKL,CT values were comparable or below the reference doses: 65 mGy and 758 mGy cm, respectively at head CT; 11.5 mGy and 327 mGy cm, respectively at chest CT; 11.6 mGy and 437 mGy cm, respectively at abdominal CT; and 11.0 mGy and 264 mGy cm, respectively at pelvis CT. Estimated effective doses were 1.6, 4.6, 6.6 and 4.0 mSv, respectively. The study offered a first national dose survey and provided a mean for quality control and optimization of CT practice within the country.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ejrad.2010.12.050Additional details
Identifiers
- DOI
- 10.1016/j.ejrad.2010.12.050;
- PII
- S0720-048X(10)00675-3;
Publishing Information
- Journal Title
- European Journal of Radiology
- Journal Volume
- 80
- Journal Issue
- 3
- Journal Page Range
- p. e544-e551
- ISSN
- 0720-048X
- CODEN
- EJRADR
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43078359
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ADULTS; CHEST; COMPUTERIZED TOMOGRAPHY; HEAD; KERMA; PATIENTS; PELVIS; PHANTOMS; QUALITY CONTROL; RADIATION DOSES; SUDAN
- Descriptors DEC
- AFRICA; AGE GROUPS; ARAB COUNTRIES; BODY; CONTROL; DEVELOPING COUNTRIES; DIAGNOSTIC TECHNIQUES; DOSES; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.