A modular electronic readout system for the OPAL electromagnetic presampler
Creators
- 1. Enrico Fermi Inst., Dept. of Physics, Univ. Chicago, IL (United States)
Description
The readout system used to sample and digitize the 21 504 channels of the OPAL electromagnetic barrel presampler is described. The system consists of semi-custom analog electronics mounted directly on the detector to provide integrate-and-hold capability and custom VME boards to implement the digitization functions. The data from one event is digitized completely in under 7 ms. This meets the OPAL requirement of a sustained 10 Hz trigger rate with less than 10% dead time. A minimum ionizing particle deposits 2.7 pC in a single layer of the presampler, and the largest signal commonly seen in 200 pC. The dynamic range of the readout system is 0.032 to 1000 pC. The noise on a single channel is typically 0.2 pC. A linear analog calibration is automatically applied on a channel-by-channel basis. Programmable logic is used extensively. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A
- Journal Volume
- 325
- Journal Issue
- 3
- Journal Page Range
- p. 494-508.
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 24036202
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANALOG SYSTEMS; BACKGROUND NOISE; CALIBRATION; CHARGED PARTICLE DETECTION; DATA ACQUISITION SYSTEMS; DEAD TIME; DIGITIZERS; LEP STORAGE RINGS; LOGIC CIRCUITS; MODULAR STRUCTURES; READOUT SYSTEMS; STREAMER SPARK CHAMBERS
- Descriptors DEC
- DETECTION; ELECTRONIC CIRCUITS; GAS TRACK DETECTORS; MEASURING INSTRUMENTS; NOISE; RADIATION DETECTION; RADIATION DETECTORS; SPARK CHAMBERS; STORAGE RINGS; TIMING PROPERTIES