Dose dependence and uncertainty evaluation of calibration factors for active neutron personal dosimeters
- 1. Korea Research Institute of Standards and Science, Daejeon (Korea, Republic of)
Description
Active neutron personal dosimeters are now widely used to monitor the neutron personal dose equivalent. They are calibrated periodically in the standard neutron fields such as 252Cf, D2O-moderated 252Cf, and 241Am-Be sources. Type B uncertainty of calibration factors should be determined if they are not offered by the manufacturer. DMC 2000 GN and EPD-N2 dosimeters were chosen to evaluate Type B uncertainties. Dose dependence of calibration factors and minimum dose to the dose indication were also investigated with 252Cf source neutrons in the neutron irradiation room at Korea Research Institute of Standards and Science. The neutron dose was irradiated up to around 10 mSv for studying dose dependence. The calibration factor of EPD-N2 was stable above neutron dose around 2 mSv and that of DMC 2000 GN was stable above around 1 mSv. Type B uncertainties of DMC 2000 GN and EPD-N2 were evaluated to be 6% and 12% from the serial measurements. Type B uncertainty of EPD-N2 was also evaluated by fitting the normalized normal distribution of calibration factors with Gaussian function from 153 dosimeters. The result was in good agreement with the result from the serial measurement. Study of minimum dose to the dose indication showed that DMC 2000 GN is more sensitive than EPD-N2. (author)
Availability note (English)
Available from http://dx.doi.org/10.15669/pnst.4.798Additional details
Identifiers
- DOI
- 10.15669/pnst.4.798;
Publishing Information
- Journal Title
- Progress in Nuclear Science and Technology
- Journal Volume
- 4
- Journal Page Range
- p. 798-801
- ISSN
- 2185-4823
Conference
- Title
- 12. international conference on radiation shielding; RPSD-2012: 17. topical meeting of the Radiation Protection and Shielding Division of ANS
- Acronym
- ICRS-12
- Dates
- 2-7 Sep 2012
- Place
- Nara (Japan)
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 49058860
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; DATA COVARIANCES; DOSE EQUIVALENTS; DOSEMETERS; GAUSS FUNCTION; NEUTRON DOSIMETRY; NEUTRON MONITORS; PERSONNEL DOSIMETRY; RADIATION DOSES; SEMICONDUCTOR DETECTORS; SENSITIVITY
- Descriptors DEC
- DOSES; DOSIMETRY; FUNCTIONS; MEASURING INSTRUMENTS; MONITORS; RADIATION DETECTORS; RADIATION MONITORS
Optional Information
- Notes
- 3 refs., 3 figs., 1 tab.