Published December 2013 | Version v1
Journal article

Development of soft X-ray energy spectrometer on Shenguang Ⅱ-up equipment

  • 1. Research Center of Laser Fusion, CAEP, Mianyang (China)

Description

Soft X-ray energy spectrometer (SXS) is an important instrument for radiation temperature measurement. According to Shenguang Ⅱ-up experimental requirements, the 15-channel soft X-ray energy spectrometer has been developed and its performances have been studied. The spectrometer's structure has been improved. Circularity convex has been used to image the target for precise collimation, and angle-changeless setup has been adopted to avoid mirrors angle error. Filters, mirrors and X-ray diodes (XRDs) have been calibrated on synchrotron radiation facility. Then the results show that SXS spectrum range is 0.05-4.97 keV. And the temporal performances have been studied on the short-pulse laser facility. The temporal resolution achieves to 99.22 ps. (authors)

Additional details

Publishing Information

Journal Title
High Power Laser and Particle Beams
Journal Volume
25
Journal Issue
12
Journal Page Range
p. 3139-3144
ISSN
1001-4322

Optional Information

Notes
8 figs., 2 tabs., 10 refs.