Published August 1988 | Version v1
Journal article

TFTR epithermal neutron detector system: Recalibration and effect of nonisotropic neutron emission

  • 1. Princeton University, Plasma Physics Laboratory, Princeton, New Jersey 08543

Description

The primary TFTR neutron source strength measurement system consists of seven fission detectors previously calibrated with D--D and D--T neutron generators and a 252Cf neutron source inside the TFTR vacuum vessel. A recalibration became desirable because of the addition of major components to the tokamak. The new calibration with the D--D neutron generator in situ is consistent with the detection efficiencies measured in the previous calibrations, within the uncertainties. Effects of the anisotropic emission of the neutron generator, due both to the variation of the differential D--D yield with angle (similar to that from beam--target and beam--beam reactions in the beam-driven TFTR plasma) and to scattering and absorption by the generator heads have been observed

Additional details

Publishing Information

Journal Title
Rev. Sci. Instrum.
Journal Volume
59
Journal Issue
8
Series
Rev. Sci. Instrum.
Journal Page Range
1682-1684
ISSN
0034-6748
CODEN