A ceric-cerous sulfate dosimeter
Creators
- 1. Academia Sinica, Beijing, BJ (China). Inst. of Atomic Energy
Description
With a spectrophotometer at wavelength of 320 nm, the molar linear absorption coefficient determined for Ce4+ ions in acidic sulfate solution is 599.1 m2/mol and its linearity is better than 0.5% in the absorption range from 0.1 to 1.6. The repeatability of the background readings for dosimeters is better than 1%. The G value of Ce3+ ions of the system is 2.39 x 10-7 mol/J(+-1.2%), and the dose conversion factor K is 724.8 Gy. From the experiment, it is found that G value will decrease 0.25% if the temperature rises 1 degree C. The measuring range for the dosimeter is from 4 to 35 kGy and the total uncertainty is 4% (95% confidence). The post-irradiation experiments of the dosimeters have been carried out. If dosimeters are kept from sunlight, they are stable in 34 days after irradiation. It is shown that the ceric-cerous sulfate can work as transfer dosimeter or standard dosimeter for measurement
Availability note (English)
MF available from INIS under the Report Number.Files
26050682.pdf
Files
(266.6 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:6b7911d82b6027f7c1776f4ce9874b29
|
266.6 kB | Preview Download |
Additional details
Publishing Information
- ISBN
- 7-5022-1209-4
- Imprint Pagination
- 8 p.
- Report number
- CNIC--00854
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 26050682
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CERIUM SULFATES; CHEMICAL DOSEMETERS; CHEMICAL REACTION YIELD; CONCENTRATION RATIO; ERRORS; INTERLABORATORY COMPARISONS; RADIATION DOSES; RESPONSE FUNCTIONS; STABILITY; TEMPERATURE DEPENDENCE
- Descriptors DEC
- CERIUM COMPOUNDS; DOSEMETERS; FUNCTIONS; MEASURING INSTRUMENTS; OXYGEN COMPOUNDS; RARE EARTH COMPOUNDS; SULFATES; SULFUR COMPOUNDS; YIELDS
Optional Information
- Secondary number(s)
- IAE--0131.