Published June 1, 2021 | Version v1
Journal article

Reduction of the effect of internal activity in LaCl3:Ce scintillator

  • 1. Saha Institute of Nuclear Physics, HBNI, 1/AF Bidhannagar, Kolkata 700 064 (India)

Description

Despite having excellent energy and time resolutions, the intrinsic α and β contaminations in the lanthanum halide scintillators pose severe limitations in their usage in rare-event detections, especially. In the present work, pulse shape discrimination (PSD) with a pre-defined algorithm of a commercially available fast digitizer has been utilized to separate the effect of α contamination from the spectrum. The efficacy of the method has been measured experimentally and compared with simulation. The contribution of the β activity has been eliminated by generating its shape with the help of a Monte-Carlo based simulation code. Thus the background events generated by intrinsic β and α activities have been reduced appreciably. The present study will encourage the application of these detectors in low cross-section measurement experiments. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/16/06/P06025

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
16
Journal Issue
06
Journal Page Range
[17 p.]
ISSN
1748-0221