Evaluation of radiation dose to pediatric patients during certain special procedures
- 1. Salman bin Abdulaziz University, College of Applied Medical Sciences, Radiology and Medical Imaging Department, Alkharj, P.O. Box 422 (Saudi Arabia)
- 2. Radiological Sciences Department, College of Applied Medical Sciences, King Saud University, P.O. Box 10219, Riyadh 11433 (Saudi Arabia)
- 3. Radiology and Medical Imaging Department, Alghad International Colleges for Applied Medical Sciences, Almadinah Almunawra (Saudi Arabia)
- 4. King Abdulaziz University, School of Medicine, Radiology Department (Saudi Arabia)
Description
This study was intended to measure pediatric entrance surface air kerma (ESAK) and effective dose during micturating cystourethrography (MCU), intravenous urography (IVU) and barium studies (barium meal, enema, and swallow) and to propose a local diagnostic reference level (DRL). ESAK was measured for patients using calibrated thermoluminescent dosimeters (TLDs, GR200A). Effective doses (E) were calculated using the National Radiological Protection Board (NRPB) software. A total of 236 special pediatric procedures were investigated. 21.7% of the sample comprised barium procedures, 18.6% were MCU procedures while 59.5% of the sample were IVU procedures. The mean ESAK measurements (mGy) were 2.1±0.8, 3.0±23 and 1.2±0.2 for barium meal, enema and swallow in the same order. The mean patient dose for IVU procedures was 12.4±8.7 mGy per procedure and the mean patient dose per MCU procedure was 5.8±7 mGy. Local DRLs were proposed for all procedures. The patient doses in this study are within the reported values, suggesting that pediatric patients are adequately protected. - Highlights: • Pediatric radiation dose has been evaluated for three of the most common fluoroscopic procedures. • Radiation doses were measured using calibrated TLD GR200A. • Pediatric patients of concern and ESAK doses showed large variations. • The patient doses in this study are within the reported studies suggesting that the pediatric patients are adequately protected
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radphyschem.2013.11.034Additional details
Identifiers
- DOI
- 10.1016/j.radphyschem.2013.11.034;
- PII
- S0969-806X(13)00642-7;
Publishing Information
- Journal Title
- Radiation Physics and Chemistry (1993)
- Journal Volume
- 104
- Journal Page Range
- p. 267-271
- ISSN
- 0969-806X
- CODEN
- RPCHDM
Conference
- Title
- 1. international conference on dosimetry and its applications
- Dates
- 23-28 Jun 2013
- Place
- Prague (Czech Republic)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46124291
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BARIUM; BIOMEDICAL RADIOGRAPHY; KERMA; N CODES; NRPB; PATIENTS; PEDIATRICS; RADIATION DOSES; THERMOLUMINESCENT DOSEMETERS; THERMOLUMINESCENT DOSIMETRY; VARIATIONS
- Descriptors DEC
- ALKALINE EARTH METALS; COMPUTER CODES; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; DOSIMETRY; ELEMENTS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; METALS; NATIONAL ORGANIZATIONS; NUCLEAR MEDICINE; RADIOLOGY; UNITED KINGDOM ORGANIZATIONS
Optional Information
- Copyright
- Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.